Contact coagulation
Coagulation with high depth upon direct contact between electrode and tissue
Power setting range: Max. 70 W at 200 Ω
Max. periodic HF output voltage: Max. 1,300 V
max
Max. output current 600 mA at 200 Ω
Crest factor 5.0
Danger of injury from excessive voltage!
Active accessories to be used with
this current must be able to withstand an HF voltage of at least 1,300 V.
It must be kept in mind that voltage is achieved even at settings
below 10, and that the active accessory
must be selected accordingly.