EasyManua.ls Logo

KLS Martin Minicutter - Page 41

KLS Martin Minicutter
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use
Electrosurgical Unit Minicutter
Revision 08 41
Spray coagulation
Coagulation current with high voltage for surface coagulation (fulguration).
Power setting range: Max. 60 W at 300
Max. periodic HF output voltage: Max. 2,000 V
max
Max. output current 450 mA at 300
Crest factor 5.0
Danger of injury from excessive voltage!
Active accessories to be used with this current must be able to withstand an HF
voltage of at least 2,000 V.
It must be kept in mind that voltage is achieved even at settings
below 10 and that the active accessory
must be selected accordingly.
Performance Diagram
HF power in W
Setting

Table of Contents

Related product manuals