LABOPORT Pumps and Systems Operation
Translation of original Operating Instructions, english, KNF 121211-121345 01/10 35
8.2. Switching Pump/System on and off
Switching pump on
The pump may not start up against overpressure during
switch-on. This also applies in operation following a brief powe
failure. If a pump starts against pressure, it may block. This
activates the thermal switch, and the pump switches off.
Make sure that no pressure is present in the lines during
switch-on.
Switch on pump with mains switch (see fig. 2/19)
Switching off the pump/removing from operation
When transferring aggressive media, flush the pump prior to
switch-off to increase the service life of the diaphragm (see
chapter 9.2.1, page 37).
Switch off pump with mains switch (see fig. 2/19).
Open pressure and suction lines to normal atmospheric pres-
sure.
Disconnect the power source.
8.3. Notices related to System Operation
8.3.1. Separators
You can move the holders for the separator vessels after
loosening the attachment screw.
8.3.2. High Performance Condenser
CAUTION
Danger of high performance condenser bursting
Make sure that the high performance
condenser’s upper gas outlet is not blocked.
A high performance condenser may be used only with a cold-
water connection or recirculated cooling system.
Adjustment of the high performance condenser altitude level
Qty Material
1 Allen key 5 mm
Tab. 23
1. Loosen the both allen screws at the foot of the high perform-
ance condenser.
2. Remove the high performance condenser’s clamp assembly in
the support mounting slot.
3. Retighten the allen screws.
Tools and material