Exiting the User Mode
5–2 USER MODE
3. Turn to change the parameter value.
4. To validate, press and hold for at least 3 seconds until "COPY" is displayed and a
beep is heard.
5. To keep the initial value, press on shortly. "ABOR" is displayed. The system returns to
the parameter/program mode.
Exiting the User Mode
To exit the user mode, press on shortly. “QUIT” is displayed until the system returns to the
operational mode. The self-test continues until it is completed.
Performing a Tube Seasoning
This procedure allows for a progressive warm-up of the X-ray tube. It must be performed at unit
installation and when replacing the tube head. It can also be performed when needed, for instance
after a long period of equipment inactivity. It lasts around three minutes.
To perform a tube seasoning, follow these steps:
1 Go to the User Mode menu and change parameter P 05 from OFF to ON.
The self-test continues. After the self-test is completed, I 02 is displayed. This message means
that the seasoning process must be started.
2 Press on .
The display blinks,. The sequence step number and the required exposure settings (kV, time)
are displayed alternatively.
3 Stand behind the generator.
4. Launch an X-ray.
WARNING
Be careful not to be exposed to ionizing radiation
When the exposure is completed, the display blinks. The cooling error code (I 01) and the
remaining time required before the next step are displayed alternatively.
When the cooling cycle is completed, the display blinks. The next step number and its
exposure settings are displayed alternatively.
_
+
_
+
_
+
_
+
_
+