Error Messages
6–4 TROUBLESHOOTING
Error Messages
The following table lists the error messages and the actions to take.
Table 2 Error Messages
Error
Message
Possible Cause Action
E01 + audible
alarm
The radiography control button was released
before the exposure end. The display indicates
the remaining exposure time. Based on this time,
you must decide whether to develop the film or
make another exposure).
Quickly press on the selection knob
to stop the alarm.
E02
The radiography control was enabled while the
unit was being powered ON.
Switch OFF the unit, then restart it. If
the problem persists, contact your
representative and stop using the
equipment.
E03-E04 Exposure time control problems
E10 to E18 kV voltage error
E20 to E24 Filament voltage error
E30
Problem with voltage to the main power supply
or to the chemical capacitor
E40 to E46
System error (problems with the power board
microprocessor)
E50 to E54
Problem with the IC bus, the connection
between the control panel and the power board.