KODAK 2100 Intraoral X-Ray System User Guide (SM700)_Ed04 7–1
Chapter 7
MAINTENANCE
This chapter describes the maintenance tasks you must perform for your KODAK 2100 Intraoral
X-ray system
Quarterly
Generator
Check that:
• The certification label is legible.
• No oil leaks.
Mechanical support
Check that:
• The wall framework is securely attached to the wall.
• All labels are legible.
• The scissor arm is stable in all positions.
Control timer unit and electrical installation
Check that:
• The symbols are legible.
• The control timer unit and the power supply cables are in good condition.
• The ground is installed correctly.
• The X-ray exposure button returns to its initial position after use.
Operation
WARNING
Be careful not to be exposed to ionizing radiation
Check that:
• The audible signal is heard and the X-ray emission indicator is visible when you launch an
X-ray (exposure time: 0.1 sec).
• The “E01” message is displayed when you launch an X-ray (exposure time:1.0 sec) and
you release the control button before the exposure end.
Control timer unit self-test
To enable the self-test, you must switch ON the KODAK 2100 system.
• The self-test starts with a simultaneous display and alarm light test.