Quick Troubleshooting
6–2 TROUBLESHOOTING
The developer is too cold. Heat the developer
The developer is too old or
diluted.
Replace with a new developer
The exposure time mode is
not correctly selected
Check your exposure settings (See the
exposure procedure).
The receptor is not
correctly positioned.
Reposition the receptor.
The unit is not correctly
installed.
Call your representative.
The X-ray emission works but
exposure is too dark.
The exposure time is too
long.
Reduce the exposure time.
The development time is
too long.
Reduce the development time (See the
manufacturer’s instructions).
The developer is too hot. Cool the developer
The developer is too
concentrated.
Adjust the concentration or change
developer.
The exposure time mode is
not correctly selected.
Check your exposure settings (See the
exposure procedure).
Malfunction Possible Cause Action