User Mode
The User mode allows you to choose the length of the cone (which is
necessary to calculate the correct emitted dose) and the type of imaging
receptor (required by local regulatory agencies).
Entering User Mode
1. Turn on the system.
The self-test is activated. At the end of the self-test, software
information is displayed (for example, F718 1.00).
2. Quickly press the selection knob on the control timer to enter the menu.
You have access to the menu when USER is displayed. The display
intermittently shows the first parameter (P 01) and the setting (for
example, ON).
3. To change from one parameter to another, turn the selection knob one
step in any direction.
Changing Parameters
To change parameters:
1. Press and hold the selection knob at least 3 seconds until the display
shows EDIT and you hear a sound.
The parameter value starts blinking.
2. Turn the selection knob to change the parameter value.
• To validate your choice, press and hold the selection knob at least
3 seconds until COPY is displayed and a noise sounds.
• To keep the initial value, press the selection knob briefly. “Abor”
appears on the display.
The system returns to the parameters/programs mode.
Exiting User Mode
To exit the User mode:
• Press the selection knob briefly.
“Quit” is displayed before the system return to operational mode.
N° Parameters
Choice
P01 Digital receptor ON / OFF
P02 Long cone ON / OFF