9/2007 SM700_K2100_03_en 6-1
6
Specifications
According to IEC Standard 601-2-7
Manufacturer
Trophy
A subsidiary of Carestream Health Inc.
4, rue F. Pelloutier - Croissy-Beaubourg
77435 Marne-la Vallée Cedex 2
France
Models
Dental X-ray diagnosis devices, class 1, type B, intermittent use
KODAK 2100-TR: equipped with tube TRX 708 from TROPHY
KODAK 2100-C: equipped with tube OCX / 65-G from CEI
Electric power supply (during exposure)
230–240 V AC (± 10%), 50 Hz, 5 A, apparent resistance 0.5 Ω
100–110–130 V AC (± 10%), 50/60 Hz, 12 A, apparent resistance 0.2 Ω
Electric power supply (no exposure)
230–240 V AC (± 10%), 50 Hz, 100 mA
100–110–130 V AC (± 10%), 50/60 Hz, 100 mA
Rated high voltage and maximum corresponding current
60 kV / 7 mA
Current/voltage combinations for a maximum output power of:
420 W, 60 kV / 7 mA
Rated power for exposure time of 0.1 sec.
•420 W
Rate of use
At 60 kV, 7 mA and 0.1 sec. and at the maximum tank temperature:
approximately one exposure every 8 sec.
Minimum value of the current/time product in the range of
conformity
0.07 mAs at 7 mA
Fixed parameters
60 kV / 7 mA
Area of conformity to IEC standard 60601-2-7 (2002)
Reproducibility of the emitted radiation: conform
Linearity of the emitted radiation: conform
Precision in radiography: conform