6-2 SM700_K2100_03_en 9/2007
Measurement conditions
kV: Indirect measurement using a kV peakmeter
mAs: Direct measurement in the circuit using a mAs-meter
Exposure time: Indirect measurement on the kV signal at 75% of the
peak value
Storage and transportation conditions
Temperature: -10°C to 60°C (14°F to 140°F)
Relative humidity: 10% to 95%
Atmospheric pressure: 700 to 1060 hPa
Dimensions and weight
Scissor arm
Equipped with gas jack specially designed for this application; proven to
function correctly after more than 400,000 cycles
Electromagnetic compatibility
KODAK 2100 Intraoral System complies with the European Directive
89/336/EEC and the IEC 60601.1.2 (2001) standard.
Classification: Group 1, Class B
Control unit:
13 x 9 x 4 cm
(5.1 x 3.5 x 1.6 in.)
0.15 kg
(0.33 lb)
Wall framework:
51.4 x 18.9 x 10.8 cm
(20.2 x 7.4 x 4.3 in.)
4.3 kg
(9.5 lb)
X-ray emitting unit:
43.8 x 22.6 x 12 cm
(17.2 x 8.9 x 4.7 in.)
4.3 kg
(9.5 lb)
Scissor arm:
87.3 x 13.3 x 6.3 cm
(34.4 x 5.2 x 2.5 in.)
9 kg
(19.8 lb)
Mobile stand
(optional):
90 x 60 x 110 cm
(35.4 x 23.6 x 43.3 in.)
40 kg
(88.2 lb.)
Floor column
(optional):
24 x 23 x 90 cm
(9.4 x 9.1 x 35.4 in.)
20 kg
(44.2 lb.)
Ceiling column:
50 x 50 x 154 cm
(19.7 x 19.7 x 60.6 in.)
12,8 kg
(28.2 lb.)