31
Setting
4. Press either or key to select
measurement speed.
MENU
MEAS
SP EE D MODE
NORMAL
F A S T
MULTI-N O R M A L
MULTI-F A S T
MANUAL
4-a-1. NORMAL or FAST setting: Select
[NORMAL] or [FAST] and press
ENTER
key.
The MENU - MEAS - SPEED MODE - IN-
ND screen appears.
This screen is used to determine whether the
ND filter built into the body is used or not.
4-a-2. Press either or key to select
[AUTO], [OFF] or [ON].
Select [ON] when the error message “OVER” appears.
When measuring objects having a wide range of luminance
under the adjusted measurement conditions (ex. γ
measurement), set IN-ND to [OFF] or [ON]. For [OFF] or [ON]
selection, select [OFF] when the upper limit luminance is 100cd/
m2 or less, and [ON] when it is 100cd/m
2
or more as a guide.
4-a-3. Press
ENTER
key.
After the setting is entered, the MENU
- MEAS - SPEED MODE screen
appears again on the LCD.
MENU
MEAS
IN ND
AUTO
ON
O F F
SP EE D MODE
MENU
MEAS
IN ND
AUTO
ON
O F F
SP EE D MODE
MENU
MEAS
SP EE D MODE
NORMAL
F A S T
MULTI-N O R M A L
MULTI-F A S T
MANUAL
4-b-1. MULTI-NORMAL or MULTI-FAST
setting:
Select [MULTI-NORMAL] or [MULTI-
FAST] and press
ENTER
key.
The MENU - MEAS - SPEED MODE -
MULTI INTEG screen appears.
This screen is used for input of the integration
time in the MULTI INTEG-NORMAL mode or
MULTI INTEG-FAST mode.
MENU
MEAS
MU LTI I NT EG
INTEG T IM E
0 1
I N - N D A U T O
SP EE D MODE