Chapter 3: Getting Started
Product Service 1-800-522-7658
14
Original instructions
Initial Certification
Prior to use, a qualified certifier should certify all biosafety cabinets. Under
normal operating conditions, the biosafety cabinet should be recertified at
least annually and when relocated or serviced. The certifier should perform
the following tests, as recommended in ANSI/NSF International Standard
Number 49 in effect when the cabinet was manufactured:
Downflow Velocity Profile Test
Inflow Velocity Test
Airflow Smoke Patterns
HEPA Filter Leak Test
Optional Canopy Alarm Test and Operation
Vibration Test *
Noise Level Test *
Lighting Intensity Test *
*These tests are user comfort related tests and may be omitted at the
user’s or certifier’s discretion.
If you have any questions regarding certification agencies or need assistance
in locating one, contact Labconco’s Product Service Department at 1-800-
522-7658 or 816-333-8811.