EasyManua.ls Logo

lancer FS 22 - Adjust Water Flow Rate & Syrup;Water Ratio

lancer FS 22
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
9. Turn on CO
2
at the source then, using a screwdriver, adjust
the high pressure regulator at the source to 110 PSI (0.758
MPA) then tighten locknut with wrench.
10. Adjust both of the low pressure regulators on the regulator
manifold to 75 PSI (0.517 MPA) then tighten locknut with
wrench.
11. Activate each valve until gas-out.
12. Plug in the remote pump deck, if not already done so, and
turn the switch to the ON position.
13. Activate each valve until the carbonator pump comes on.
Release the button, allow carbonator to ll and stop. Repeat
this process until a steady ow of carbonated water is
achieved.
14. Activate each valve to purge air from the syrup lines.
Adjust Water Flow Rate & Syrup/Water Ratio
1. If necessary, rotate switches panel forward and down by
releasing the two pin latches on its sides.
2. Rotate light panel, forward and up by releasing the two pin
latches on its sides towards the top.
Thewaterowcanbeadjustedbetween3.25oz/sec(96
ml/sec)and4.50oz/sec(133ml/sec)onalldispensing
valves. Ensure there is ice on the cold plate and the
linesarecoldbeforeattemptingtosettheowrateson
thevalves.Thedrinktemperatureshouldbenohigher
than40°F(4.4°C)whenowratesareset.
NOTE
There can be up to six gray water valves on this
dispenser (up to four carbonated water valves and two
plain water valves)
NOTE
4. Repeat procedure for each water valve.
A
B
C
D
Increase Decrease
A. Flow Control
B. Valve Retainer
C. Solenoid
D. Valve Body
A
B
C
D
Increase Decrease
A. Flow Control
B. Valve Retainer
C. Solenoid
D. Valve Body
5. Remove nozzle by twisting counter clockwise and pulling
down.
6. Install Lancer syrup separator (PN 82-3458) in place of
nozzle.
7. Activate valve to purge syrup until steady ow is achieved.
8. Using a Lancer ratio cup, activate the valve and capture a
sample. Verify that the syrup level is even with the water
level. Use a screwdriver to adjust if needed.
9. Remove syrup separator and reinstall nozzle.
10. Repeat Steps 5-9 for each syrup valve.
11. Re-install merchandiser and top cover.
A. Regulator Adjustment Screw
B. Adjust to 110 PSI (0.758 MPA)
C. Wrench
A
B
C
Thepumpdeckhasa3minutetimeoutfeature.Ifthe
timeoutoccurs,turnthedeckOFFthenONbyipping
the switch on the control box.
NOTE
TocheckforCO
2
leaks,closethevalveontheCO
2
cylinder and observe if the pressure to the system
dropswiththecylindervalveclosedforveminutes.
Openthecylindervalveaftercheck.
NOTE
3. Using a Lancer ratio cup verify water ow rate. Use a
screwdriver to adjust if needed.

Related product manuals