Special Conditions of Use
Equipment has the facilities to connect a charger in the non-hazardous area which has been
assessed for a maximum Um of 12V. This Um voltage is required to be maintained in
accordance with Clause 16.2 of IEC 60079-14.
Equipment is intended to be carried within a backpack in normal use. It shall be ensured that
this backpack complies with the electrostatic requirements of IEC 60079-0 to prevent
electrostatic charging of the equipment contained.
The following metal parts have been considered isolated metal parts and have the potential to
hold charge.
Main Enclosure: 1.78nF
Probe head: 18.54pF
Refer to the “Electrostatic Requirements” section of the instruction manual for details on the
mitigation of electrical discharge.
The analyzer must be carried within a backpack which suitably protects the product from
unwanted metal to metal impact and thus prevents possibility for dangerous sparking in
explosive atmospheres.
Equipment contains an FID housing certified to protection concept Ex db. Per the requirements
of this standard no modification or repair is permitted to be made to the equipment flamepaths.