34
Version 2.0, Revision I
Instrument Setup
4
Fig. 25
Fig. 26
1
2
3
Fig. 27
Filling capacities:
Each magazine (→ "Fig. 27-1") can hold up to 150 slides (depending
on individual slide thickness).
The maximum filling height of a magazine is indicated by the "max."
mark (→ "Fig. 27-2") and may not be exceeded.
Positively charged slides stick to each other so much that they
require special and careful handling to ensure reliable mechanical
handling of this type of specimen slide by the printer.
If you are using positively charged slides, please follow the
recommendations below:
• Positively charged specimen slides must be separated from each
other before they are loaded into the magazine.
• The magazine may be loaded with no more than 72 specimen slides
to prevent problems during ejection. This corresponds to the lowest
mark (→ "Fig. 27-3").
• Even smaller loads will significantly improve the ability of the
instrument to eject positively charged slides from the magazine.
4.8 Electrical connection
Warning
The instrument must be connected to a grounded mains socket.
Be sure to use the cord that is appropriate for the local power supply (plug must fit on-site wall
outlet).