Version 2.0, Revision I
Troubleshooting
7
Display
code
Source of error Troubleshooting solution
• Cover ( "Fig. 2-5") not closed properly.
Safety switch triggered.
• Flash standby state not reached within
the prescribed time. Charging electronics
defective.
• Ensure that the cover is not blocked,
e.g. by the manual unload station. Close
cover completely.
• Call Leica Technical Service.
No flash or flash duration too short.
Flash bulb dirty or defective.
Check whether flash bulb works – to do
so, observe the stray light on the cover.
Never open the cover to check whether a
flash is triggered!
No flash --> install a new flash bulb (
p. 30 – 4.6 Installing/exchanging the
flash bulb).
Unload station mechanically blocked. Remove the cause of the blockage.
Heater fan not running or running too
slowly.
Call Leica Technical Service.
No slide is in the drying module for
flashing. Slide left the specimen slide
carrier, but did not reach the drying
module.
Remove slide.
Slide not successfully ejected from drying
module.
Remove slide.
Slide in the drying module during
initialization or prior to a print job being
carried out.
Remove the cause of the blockage.
The flap of the drying module either does
not close or does not open. The flap is
blocked (e.g. by a slide).
Remove the cause of the blockage.
Lifting table end sensor does not switch. Call Leica Technical Service.
Lifting table position sensor does not
switch during positioning.
Call Leica Technical Service.
Flash power supply: Overtemperature for
more than 10 minutes.
Call Leica Technical Service.