71
Leica IP S
Troubleshooting 7
Display
code
Source of error Troubleshooting solution
Faulty control data received (program bug).
Settings for the serial interface incorrect or
the instrument configuration conflicts with
the PC configuration.
Carry out a RESET on the printer.
Check cable connection to the PC.
Check configuration of serial port of the
PC and reboot the PC.
Transmitted data do not contain
confirmation of receipt or data
transmission was not confirmed by the PC.
Follow the same procedure as for
ERROR 60.
Try another printer cable.
Print image exceeding the vertical limit. Error caused by application software.
Print image exceeding the horizontal limit. Error caused by application software.
The CRC test of the EEPROM returned an
error when the instrument was switched
on.
Call Leica Technical Service.
Internal firmware error or defective
controller.
Call Leica Technical Service.
Mechanical problem makes ejection of
slides from magazine difficult.
Check ejection mechanism. Remove
foreign bodies, then clean with brush.
A print job was attempted with a storage
fluid cartridge inserted.
Remove storage cartridge. Insert ink
cartridge and press LOADED to confirm
( p. 36 – 4.9 Exchanging the transport
cartridge for an ink cartridge).
Power supply unit does not achieve
standard operating voltage.
Call Leica Technical Service.
Firmware only partially loaded or not
loaded at all.
Flash memory defective.
Call Leica Technical Service.
Wrong firmware. Call Leica Technical Service.