EasyManua.ls Logo

Levitronix BPS-4H.4 - Maximum Static Pressure for Pump Head

Default Icon
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual for BPS-4H (High Temp.)
www.levitronix.com
PL-2009-03, Rev04, DCO# 21-101
10
Figure 9: Pressure/flow rate chart for water at 40°C
(Typical curves measured with pump head CP-4.25.)
2.6 Maximum Static Pressure for Pump Head
Figure 10: Specification for maximum static pressure of pump heads CP-4.25/CP-4.34/CP-4.36
(Pressure limits only valid for pump head mounted on motor. See also note under Section 6.1)
0
10
20
30
40
50
60
0 2 4 6 8 10 12 14 16 18
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
[psi]
[gallons/min]
[bar]
[liters/min]
Specific gravity = 1 g/cm
3
Viscosity = 0.7 cP
Liquid Temp.: 40 C
Liquid: H
2
O
4000 rpm
3000 rpm
5000 rpm
5500 rpm
6000 rpm
6500 rpm
0
20
40
60
80
100
120
77 107 137 167 197 227 257 287 317 347 377 407 437
0
1
2
3
4
5
6
7
8
9
25 45 65 85 105 125 145 165 185 205 225
[psi]
F]
[bar]
Liquid Temperature in [°C]
Max. Static Pressure for CP-4.34/36
Max. Static Pressure for CP-4.25
Safety Factor = 1.3 (according to EN12162)
Limitation: Given by deformation. Burst pressure is significantly higher.
Ambient Temp = 25 C
Note: Pressure limit only valid for pump head mounted on motor