4.5 Mechanical Installation of Adaptor Cables
For connecting the motor to the controller the adaptor cables MCAP-4.x (for power cable) and MCAS-3.x (for
sensor cable) shall be used (see Table 3 for adaptor cables). For the cables which use an M23 threaded
metallic Hummel connector type, check the connection according to the following pictures:
Figure 29: Wrong and correct Hummel connector type assembly
O-Ring visible
Wrong Connection!
O-Ring not visible
Correct Connection!
Gap >2 mm
Sensor Cable
of Motor
Power Cable
of Motor
Gap >1 mm
Gap
1 mm
Gap
2 mm
Sensor
Adaptor Cable
Power
Adaptor Cable