User Manual for BPS-i100
www.levitronix.com
PL-4036-00, Rev03, DCO# 21-057
10
2.4 Pressure-Flow Curves
Figure 11: Pressure/flow rate curves
(Measured with pump head LPP-200.7)
2.5 Maximum Static Pressure for Pump Heads
Figure 12: Specification for maximum static pressure of pump head LPP-200.7
(Pressure limits only valid for pump head mounted on motor)
0.0 1.0 2.0 3.0 4.0 5.0
0
5
10
15
20
25
30
0.00
0.25
0.50
0.75
1.00
1.25
1.50
1.75
2.00
2.25
0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0
[gallons/min]
[psi]
[bar]
[liters/min]
9000 rpm
4000 rpm
Specific gravity = 1 g/cm
3
Viscosity = 0.7 cP
Liquid Temp.: 40°C
8000 rpm
7000 rpm
6000 rpm
5000 rpm
0
20
40
60
80
100
77 87 97 107 117 127 137 147 157 167 177 187
0
1
2
3
4
5
6
7
8
25 30 35 40 45 50 55 60 65 70 75 80 85 90
[psi]
F]
[bar]
Liquid Temperature in [°C]
Max. Operational Pressure for LPP-200.7
Max. Static Pressure for LPP-200.7 (PFA Housing, PP Reinforcing Cup with SS Screws)
Safety Factor > 1.3 (according to EN12162)
Note: Burst pressure is significantly higher.
Ambient Temp = 25 °C
Note: Pressure limit only valid for pump head mounted on motor.