Operation
GA09420_002_C0 – (2016-10) –
Leybold
35
Connect the + lead of a voltmeter to the threshold measurement point of the
selected switching function ("Setpoint A" Pin 3, "Setpoint B" Pin 6) and
its lead to Pin 5.
Using a screwdriver (max. ø2.5 mm), set the voltage of the selected
switching function (Setpoint A, B) to the desired value U
Threshold
.
Setting of the switching functions is now concluded.
There is no local visual indication of the statuses of the switching func-
tions. However, a functional check of the switching functions (On/Off)
can be made with one of the following methods:
Reading the status via fieldbus interface [3].
Measurement of the relay contacts at the sensor cable connector
with a ohmmeter/continuity checker ( 20).
max. ø2.5
Setpoint
A
Pin 3
Setpoint
B
Pin 6
Pin 5