On-Screen Messages
Messages Explained
Message Explanation
<1.3L Retest or Try Manual Test Breath fl ow ended before the subject blew
1.3 liters of breath. Instruct subject to try
again or use manual test mode.
>0.6 BAC is unusually high and above 0.6 BAC.
Subject may require medical attention.
External Interference External interference has been detected.
Move to a different location and try again.
Flow Error - Retry & Blow Steadily Exhalation not complete or interrupted.
Instruct subject to blow steadily as long as
they can.
Invalid Calibration/Cal Check No alcohol was detected. Repeat setup,
then retest.
Log Empty There are no results in the memory.
Low Battery Battery voltage is too low to take a test.
Replace batteries.
Pump Reset Needed The pump needs to be reset. Follow
onscreen instructions.
Temperature When calibrating or taking a test, unit is
outside of temperature limits.
Calibration range = 68
o
- 95
o
F (20
o
- 35
o
C)
Testing range = 32
o
- 130
o
F (0
o
- 55
o
C)
Timeout User blew for > 15 seconds.
Please try again, blow harder.