Maintenance
Valid for all LINAK products
LINAK
®
products must be cleaned at regular intervals
Inspect for malfunction, mechanical damage, wear and cracks. Worn-out parts must be replaced
Inspection/maintenance intervals may be defined by the medical device manufacturer
LINAK products are closed units and require no internal maintenance
Only type IPX6, IPX6 Washable and IPX6 Washable DURA
TM
are waterproof
LINAK products must be IPX6 Washable and IPX6 Washable DURA
TM
when cleaning in wash tunnels
O-rings: When individual parts are replaced in a LINAK IPX6, IPX6 Washable or IPX6 Washable DURA
TM
system, the O-rings must be replaced at the
same time on all parts
On all products where replaceable cables or fuses have been dismounted or replaced, the O-ring must be replaced, and the O-rings and the
receptacle insert must be greased with an acid-free Vaseline.
Valid for all LINAK actuators and lifting columns
Actuators/lifting columns must be inspected at attachment points, wires, piston rod, enclosure, and plugs, and it must be checked that the
actuators/lifting columns function correctly
To ensure that the pregreased inner tube remain lubricated, the actuator must only be washed when the piston rod is fully retracted
Valid for all LINAK control boxes and hand controls
Electronics must be inspected at attachment points, wires, enclosure, and plugs
Inspect the connections, cables, enclosure, and plugs, and check for correct functioning
The control box is sealed and maintenance-free
Environmental conditions
Operating, storage and transport
Operating
Temperature
Relative humidity
Atmospheric pressure
5 °C to 40 °C
20% to 80% - non-condensing
700 to 1060 hPa (Rated to be operated at an altitude ≤ 3000 m)
Storage
Temperature
Relative humidity
Atmospheric pressure
-10 °C to +50 °C
20% to 80% - non-condensing
700 to 1060 hPa (Rated to be stored at an altitude ≤ 3000 m)
Transport
Temperature
Relative humidity
Atmospheric pressure
-10 °C to +50 °C
20% to 80% - non-condensing
700 to 1060 hPa (Rated to be transported at an altitude ≤ 3000 m)
If the actuator is assembled in the application and is exposed to push or pull during transportation, the actuator can be damaged.
Do not drop an actuator or otherwise damage the housing during disassembly or transportation.
We do not recommend to use an actuator which has been damaged.
Valid if nothing otherwise is stated under the specific products in a later section.