6.3 Electrical Specications (only Sprint 200 with i-Drive Power)
Parameter Value
Input Voltage, Frequency (Sprint 200 with i-Drive Power) 100V AC, 50/60 Hz
110V AC, 50/60 Hz
120V AC, 50/60 Hz
127V AC, 50/60 Hz
230V AC, 50/60 Hz
Maximum Power Input (Sprint 200 with i-Drive Power) 400 VA
Ingress Protection according to EN 60529
(Sprint 200 with i-Drive Power)
IPX4
Batteries of the i-Drive Power system 3 x 12V 9Ah VRLA
Fuses in the i-Drive Power system
Version 100V AC
Version 110V AC
Version 120V AC
Version 127V AC
Version 230V AC
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T1,6A L 250V
Electric Protection Class (Sprint 200 with i-Drive Power) Class I
Parameter Value
External Dimensions in Standard Stretcher Position (length x width) 216 cm x 89 cm
Maximum Siderail Height above Mattress Support Platform 40 cm
Siderail Length (Side Protection Zone for Patient) 137 cm
Distance between siderail bars 20,6 cm
Mattress dimensions (length x width) 203 cm x 76 cm
Castor diameter 20 cm
FlexiDrive Castor diameter 16 cm
i-Drive Power Wheel diameter 21 cm
Clearance of Undercarriage in Standard Position 10,7 cm
Minimum — Maximum Mattress Support Platform Height above oor (wit-
hout Mattress)
53 cm — 86 cm
Maximum Backrest Angle 90°
Maximum Thighrest Angle 40°
Maximum Calfrest Angle 25°
Trendelenburg Tilt Angle / Anti-Trendelenburg Tilt Angle +17°/-17°
Maximum Angle between Calfrest and Thighrest 115°
Average Stretcher Weight of the Sprint 200 without i-Drive Power 143 kg / 315 lb
Average Stretcher Weight of the Sprint 200 with i-Drive Power 154 kg / 340 lb
SWL (Stretcher Safe Working Load) 320 kg / 705 lb
Maximum Patient Weight 280 kg / 617 lb
Maximum Mass of Mobile Hospital Bed (Maximum Mass of Empty Stret-
cher without i-Drive Power + Safe Working Load)
486 kg / 1071 lb
Maximum Mass of Mobile Hospital Bed (Maximum Mass of Empty Stret-
cher with i-Drive Power + Safe Working Load)
497 kg / 1096 lb
Ergoframe Distances (Backrest Distance/Thighrest Distance) 7,5 cm / 3 cm
Maximum Sound Pressure Level 71 dBA

Table of Contents

Other manuals for LINET Sprint 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Sprint 200 and is the answer not in the manual?

LINET Sprint 200 Specifications

General IconGeneral
BrandLINET
ModelSprint 200
CategoryMedical Equipment
LanguageEnglish

Related product manuals