Gocator Line Profile Sensors: User Manual
Gocator Device Files • 399
Element Type Description
Measurements\MaxErrorZ Line tool
measurement
Maximum Error in Z measurement.
Features\Line
GeometricFeature
Line LineFeature.
Features\ErrorMinPoint GeometricFeature ErrorMinPoint PointFeature.
Features\ErrorMaxPoint GeometricFeature ErrorMaxPoint PointFeature.
Element Type Description
@id 32s Measurement ID. Optional (measurement disabled if not
set).
Name String Measurement name.
Enabled Boolean
Measurement enable state:
0 – Disable
1 – Enable
HoldEnabled Boolean
Output hold enable state:
0 – Disable
1 – Enable
SmoothingEnabled Boolean
Smoothing enable state:
0 – Disable
1 – Enable
PreserveInvalidsEnabled Boolean Preserve invalid measurements enable state
0 – Disable
1 – Enable
SmoothingWindow 32u Smoothing window.
Scale 64f Output scaling factor.
Offset 64f Output offset factor.
DecisionMin 64f Minimum decision threshold.
DecisionMax 64f Maximum decision threshold.
Percent
(Percentile measurement only)
64f Error percentile.
Line Tool Measurement
ProfilePanel
A ProfilePanelelement defines settings for a profile panel tool and one or more of its measurements.
Element Type Description
Name String Tool name.
Features
Collection Not used.
ProfilePanel Child Elements