EasyManua.ls Logo

LogTag UTRED-16 - Stopping the UTRED-16; Automatic Stop; Manual Stop; Logger Stopped Status

LogTag UTRED-16
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating the Logger UTRED-16 User Guide 33
Stopping the UTRED-16
Automatically
The UTRED-16 automatically stops recording temperature when the maximum number of
readings specified during configuration has been reached. Your unit can also be set up to
stop when it is inserted into a USB socket. This option needs to be set up at the factory
and cannot be changed during configuration with LogTag
®
Analyzer. Your distributor can
supply more information about this option.
Manually
You can configure a UTRED-16 so it can be stopped with the button. This feature is useful
when you take the logger out of a shipment and don’t want to falsify the statistics with
readings taken after the shipment completion. The stop function is enabled in the
Advanced Options dialogue during configuration.
Press and hold the button. The STOPPED symbol shows in
addition to for 2 seconds.
The symbol turns off after 2 seconds. Release the
button and the logger stops taking temperature readings.
If the button is released while is still on, or you wait until STOPPED disappears, the
display shows again, and the logger continues to record data. In this case, an
inspection mark will be recorded in the data against the next temperature reading. The
MARK symbol will show, and the PAUSED symbol also shows if this feature is enabled.
Unlike many other LogTag
®
products with display, the alarm does not have to be
cleared before the logger can be stopped.
Logger has stopped
Once the logger has stopped, the display will show:
l the STOPPED symbol
l the OK indicator ( ) if no alarm was recorded
l the alarm indicator ( ) and one or both alarm threshold( ) indicators if an alarm
was recorded during the trip that has not been cleared
l the current time
l the battery low symbol ( ) if the battery is low
The logger does not show a temperature reading.