Touchclave Ecotech
21
This is used on certain loads that require drying like fabrics and pipette tips and is known as the “post-
vacuum” stage.
Printer (optional)
A thermal printer is used to print cycle data and so creating a hard copy of the cycle data.
Water Tanks (optional)
This option includes a water fill tank and a drain tank, preventing the need for a mains water supply
and external drain. The autoclave controller will instruct the user when the water tank requires filling
and the drain tank requires emptying ensuring that the autoclave can be used in almost any lab
environment.
USB Data Archiving
With this option, all cycles are recorded onto an integral micro SD card. The record of each cycle
shows the date, time, batch number, operator identification and all cycle data or failures. The record
can be printed via thermal printer (if fitted), or transferred onto a USB stick using the USB port on the
front panel of the machine.