Section 6- Alarms
All the System alarms are detailed in this section with a brief description of what each alarm means.
For troubleshooting of alarms see Service Manual.
Purge Timeout
Steam purge stage has failed to achieve the load and vent probe temperatures required to move on
to the next stage of the cycle in the time period set in the controller software.
Heatup Timeout
Heatup stage has failed to achieve the vent and load probe temperatures required to move on to the
next stage of the cycle in the time period set in the controller software.
Stabilisation Timeout
Stabilisation stage has failed due to vent and load temperature probe temperatures not remaining in
the band set in software for a time period set in controller software.
Cooling Timeout
Cooling stage has failed to achieve the load and vent temperature probe temperatures to move on
the next stage of the cycle in the time period set in the controller software.
Under Temp
Either the vent or load probe temperature has fallen below the sterilisation temperature.
Over Temp
Either the Vent or Load probe temperature has gone above the sterilisation temperature.
Power Failure
Power was lost to the machine while in the middle of a cycle.
Chamber Transducer Fail
Chamber pressure transducer is not reading correctly.
Steam Gen Transducer Fail
Steam generator transducer is not reading correctly.
Vent Probe Fail
Vent temperature Probe is not reading correctly.