9 Troubleshooting
Before sending the pyrometer for repair, try to find the error and to solve the problem with the
help of the following list.
Temperature indication too low
• Incorrect alignment of the pyrometer to the object
⇒ New correct alignment to achieve the max. temperature signal
• Measuring object smaller than spot size
⇒ check measuring distance, smallest spot size is at nominal measuring distance
• Measuring object is not always in the measuring spot of the pyrometer
⇒ Use max. value storage (see 4.9
)
• Emissivity set too high
⇒ Set lower correct emissivity corresponding to the material (see 4.5
)
• Lens contaminated
⇒ Clean lens carefully (see 6.2
)
Temperature indication too high
• Emissivity set too low
⇒ Set lower correct emissivity corresponding to the material (see 4.5
)
• The measurement is influenced by reflections of hot machine parts
⇒ Use mechanical construction to avoid the influence of the interfering radiation
Measuring errors
• Indicated temperature is decreasing during the use of the pyrometer, contamination of the
lens
⇒ Clean lens. Recommendation: use of air purge (see 3.3
)
• Air contamination in the sighting path between pyrometer and object
⇒ Change position of the pyrometer with a clean sighting path (if necessary use a ratio
pyrometer)
• HF-interferences
⇒ Correct the connection of the cable shield (see 3.1
)
• Temperature Indication is fluctuating, probably caused by changing emissivity
⇒ Wrong pyrometer type, use of ratio pyrometer recommended
LumaSense Technologies Temperature and Gas Sensing Solutions
上海麦兴 版权所有
!! !
麦兴(中国)有限公司
地址:香港湾仔告士打道151号国卫中心11楼
电话:
传真:
邮箱:
(852) 2836 8361
(852) 3011 5863
mx@imaxsun.com
www.imaxsun.com
上海麦兴仪器设备有限公司
地址:上海市浦东新区张杨路188号汤臣中心A座
电话:
传真:
邮箱:
邮编:
(86 21) 5888 7876
mx@imaxsun.com
MaxSun (China) Limited.
(86 21) 5888 6718 / 133 8186 8102
200122
Shanghai MaxSun Industrial Co., Ltd.