Marquette Hellige GmbH MicroSmart V 2.xx Page 44
Servicing Instructions 227 470 35 B - 97.12
5 Selecting the type of appliance
5.1 General
The unit provides functions for selecting the type of appliance. The type of appliance defines
certain performance features of the unit. The following types of appliances are available:
Type: International After power-up the unit displays "MicroSmart". The company name
"marquette HELLIGE" is printed on the recordings. The electrodes
are labeled "R, L, F, N, C1, C2, C3, C4, C5, C6". The date format
is "dd.mm.yy". The following languages can be selected in the
configuration:
German, English, French, Italian, Spanish, Portuguese, Swedish,
Russian. The line filter is set for 50 Hz.
Type: USA Keyboard with text instead of icons. Company name "marquette"
appearing on the unit and on the type plate. The company name
"marquette" is printed on the recordings. A place holder for Name
instead of the name of the unit is printed on the real-time
recordings (Man/Arrhy). Default (works) setting the notation
(electrode labels) is AAMI. The date format is "mm.dd.yy".
Languages: English, français, español, português. No data
communication, i.e. the following items are not displayed in the
configuration: copy function, data transfer, tel. No. Default (works)
setting for override function is On. The line filter is set for 60 Hz.
Default (works) setting for paper format is Roll (cannot be altered).
The time format is "am" and "pm". Default (works) setting for the
recording format is: sequential, rhythm record Yes. Lead group
definitions:
Stnd (standard) in place of Eint (Einthoven)
Aug (augmented) in place of Gold (Goldberger)
V1-3 (chest V1-3) in place of Wil1 (Wilson 1)
V4-6 (chest V4-6) in place of Wil2 (Wilson 2)
Program designation on recording and on display:
12Ld in place of Auto
3Ld
in place of Man
HR^
in place of Arrh
Type: USA2 Device name MAC 500 in place of MicroSmart. Otherwise as type
"USA"
Type: Asia Company name "marquette" appearing on the unit and on the type
plate. The company name "marquette" is printed on the recordings.
The following languages can be selected in the configuration: