EasyManua.ls Logo

marquette hellige MicroSmart - Patient Leakage Current Test

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Marquette Hellige GmbH MicroSmart V 2.xx Page 57
Servicing Instructions 227 470 35 B - 97.12
Test has failed if the measured values are greater than:
N.C. S.F.C
100 µA 500 µA
300 µA (U.L. requirements)
Electrical diagram for Enclosure Leakage Current Test
8.2.5 Patient Leakage Current Test
This test performs a leakage current test under single fault conditions (S.F.C.) dependent of
domestic power outlet with 115 or 230 V AC as source into the floating inputs.
In any case, the leakage current is measured from Input Jack, of unit under test, to ground.
For testing the ECG input, a patient cable, with all leads connected together, is used.
Connect the unit under test to your Safety Tester.
- Referring to the electrical diagram, measurements have to be done under following
conditions
* Polarity switch NORM and RVS
* GND switch GND closed
*S1 closed
Test has failed if the measured values are greater than 50 µA

Table of Contents