Marquette Hellige GmbH MicroSmart V 2.xx Page 48
Servicing Instructions 227 470 35 B - 97.12
Set the contrast with new display
Replacing the power supply
First remove the PCB MicroSmart (see above). Detach the four attachment screws on the metal
carrier base (power supply module). Take out the power supply module. The power supply is
bolted to the metal carrier base with 3 attachment screws. Disconnect the plug connections
CB
and CA. Detach the three attachment screws. Replace the power supply.
Replacing the patient input
First remove the PCB MicroSmart (see above). Detach the four attachment screws on the metal
carrier base (power supply module). Take out the power supply module. The patient input is
bolted to the metal carrier base with 2 attachment screws. Detach the attachment screws.