Marquette Hellige GmbH MicroSmart V 2.xx Page 53
Servicing Instructions 227 470 35 B - 97.12
Analyzing the ECG signal and the HR value
Select the following settings at the ECG simulator:
- amplitude 1 mV
- heart rate (RATE) 60 P/min
Connect the electrode leads in the following assignment:
R red -----> RA
L yellow -----> LA
F green -----> LL
N black -----> RL
C1 white/red -----> V1
: : :
C6 white/purple -----> V6
Switch to the
"Man" mode and start recording by pressing the Start/Stop key. Check by pressing
the lead stepping keys if all leads
are registered.
The ECG waveforms must be free of noise.
Record two pages in the "Man" mode. The following lettering must be visible:
- text in the line in the top margin:
first sheet
Device name
Software Version
second sheet
Company name
Date
Time
- text in the line in the bottom margin:
first sheet
error messages
operating mode
speed
sensitivity
ADS
second sheet
error messages
line filter
frequency range
heart rate