EasyManua.ls Logo

Masimo EMMA - Electromagnetic Compatibility; Electromagnetic Emissions Guidance

Masimo EMMA
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.masimo.com 48 Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Emissions
EMMA is expected to be used in a professional environment like intensive care unit, patient room and operating suite
except near high frequency electro surgical equipment and magnetic resonance imaging (MRI) systems. EMMA is also
intended to be used in emergency medical services as road ambulances.
EMMA is intended for use in the electromagnetic environment specified in the tables below. Customers or end users of
EMMA should assure that the EMMA is used within the intended environment.
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
Emission
Test
Compliance Electromagnetic Environment - Guidance
RF
Emissions
CISPR 11
Group 1
ME Equipment uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF
Emissions
CISPR 11
Class B
EUROCAE ED-14G or RTCA DO
160G, Section 21, category M
levels
Suitable for use in all professional facility and home healthcare
environments.
Suitable for use in air ambulances.

Other manuals for Masimo EMMA

Related product manuals