Guidance and Manufacturer's Declaration-Electromagnetic Immunity
This section constitutes the guidance and Masimo’s declaration regarding electromagnetic immunity for EMMA.
Essential performance of EMMA is gas measurement accuracy (see Citation [4]) including gas reading alarm
conditions, or generation of technical alarm conditions, when exposed to the following immunity levels.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test Compliance Level Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/-8 kV contact
+/-2 kV, +/-4 kV, +/-8 kV,
+/-15 kV air
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Power frequency (50
/ 60 Hz) magnetic
field.
IEC 61000-4-8
30 A/m Power frequency magnetic fields should be at levels
characteristic of typical location in a typical hospital
environment.
Radiated RF
IEC 61000-4-3
10 V/m
80%AM@1kHz
80 MHz to 2.7 GHz and
Table 9 (60601-1-2:2015)
Portable and mobile RF communications equipment should be
used no closer than 30 cm to any part of EMMA. Otherwise,
degradation of the performance of this equipment could result.