NomoLine ISA Service Manual Chapter 2: Maintenance Procedures
www.masimo.com 30 Masimo
Troubleshooting:
Check that the Nomo extension’s Luer connector is tightly connected to the gas
regulator.
Check that the float of the rotameter lies above the first line during the entire test
and that you allow a minimum of 30 seconds of stabilization before noting the gas
values.
Check the expiration date and the storage condition for the gas cylinder. Also
make sure that the cylinder´s gas pressure is sufficient, i.e. the pressure indicator
on the manometer is not within the red mark (see figure 1-16).
Leave the NomoLine ISA system running with the NomoLine ISA Maintenance Kit
connected for a minimum of 15 minutes before running the gas span check again.