NomoLine ISA Service Manual Chapter 3: Troubleshooting
www.masimo.com 32 Masimo
Symptom Possible Cause Correction
Gas readings are
questionable
Incorrect sampling line
used
Sampling line attached
incorrectly
Sampling line leaking
Internal error
Use the correct sampling
line with the NomoLine ISA
device.
Disconnect and reconnect
the sampling line to the
NomoLine ISA device.
Follow the Directions for Use
to properly connect the
sampling line.
Check sampling line for
leaks.
Replace sampling line.
If persistent, the NomoLine
ISA device requires service.
See Return Procedure on
page 35.