18. PRM-8000 Speci cations¹
Battery Life²
Standard 9V Lithium
60,000 Hours / 6.85 Years
Standard 9V Alkaline
28,250 Hours / 3.22 Years
Measurement
Front End
Detector
LND 712 Geiger-Müller tube, Halogen-quenched with mica
end window
Radiation Detected
alpha, beta, gamma and x-rays
Energy Sensitivity
alpha: minimum 2.5 MeV; ef ciency at 3.6 MeV > 80%
beta: minimum 50 keV with 35% ef ciency at 150 keV 75%
gamma/X rays: min 10 keV (window), min 40 keV (case)
1000 CPM/mR/hr
Calibrated to Cesium-137 (Cs137)
Accuracy
±10% Typical, ±15% Maximum
Measurement Ranges
0.001 to 200 mR/hr
0.01 to 2000 uSv/hr
0 to 3,333 CPS (16.66 CPS/mR/hr)
0 to 200,000 CPM (1000 CPM/mR/hr)
Total Count: 0 to 4,294,967,295 (32 Bit Integer)
Measurement Statistics
Continuous availability of Average, Minimum and Maximum
dose rate measurements in
either uR/hr, mR/hr, uSv/hr, CPS or CPM
Time stamp on new maximum
Dose Rate Alarm
Alarm can be set in mR/hr or uSv/hr to any value within the
measurement range.
Internal Data Logging
Storage
100,000 Bytes non-volatile memory
Logging Intervals
Every Minute or Every Hour
Maximum Capacity
91,466 Minutes or Hours
(90% Normal Background, 10% Full-Scale)
Minimum Capacity
32,656 Minutes or Hours
(100% Full-Scale Measurement Levels)
Logging Output Format
Standard Comma Separated Value (CSV) with time stamp
and CPM Data (1000 CPM/mR/hr)