40 | PRM-7000 / 8000 / 9000 Users Guide
19. PRM-9000 Speci cations¹
Battery Life²
Standard 9V Lithium
42,850 Hours / 4.89 Years
Standard 9V Alkaline
20,150 Hours / 2.30 Years
Measurement
Front End
Detector
LND 7317 Pancake Geiger-Muller tube, Halogen-quenched
with mica window and 45mm (1.75 inch) effective diameter
Radiation Detected
alpha, beta, gamma and x-rays
Energy Sensitivity
alpha: minimum 2.0 MeV
beta: minimum 0.16 MeV with 25% ef ciency at 1 MeV
gamma/X rays: min 10 keV (window)
gamma sensitivity: 3,500 CPM/mR/hr
calibrated to Cesium-137 (Cs137)
Accuracy
±10% Typical, ±15% Maximum
Measurement Ranges
0.001 to 125 mR/hr
0.01 to 1,250 uSv/hr
1 to 125,000 uR/hr
1 to 437,500 CPM (3,500 CPM/mR/hr)
0 to 7,291 CPS (58.33 CPS/mR/hr)
Total Count: 0 to 4,294,967,295 (32 Bit Integer)
Measurement Statistics
Continuous availability of Average, Minimum and Maximum
dose rate measurements in
either uR/hr, mR/hr, uSv/hr, CPS or CPM
Time stamp on new maximum
Dose Rate Alarm
Alarm can be set in mR/hr or uSv/hr to any value within the
measurement range.
Internal Data Logging
Storage
100,000 Bytes non-volatile memory
Logging Intervals
Every Minute or Every Hour
Maximum Capacity
91,466 Minutes or Hours
(90% Normal Background, 10% Full-Scale)
Minimum Capacity
32,656 Minutes or Hours
(100% Full-Scale Measurement Levels)
Logging Output Format
Standard Comma Separated Value (CSV) with time stamp
and CPM Data (1000 CPM/mR/hr)