Meter Specifications
Result Range: 20 - 600 mg/dL
Sample Size: Minimum 0.5 microliters (0.5 µL)
Sample Type: Fresh capillary whole blood from the finger or forearm,
venous blood drawn in EDTA (purple top tube) or heparin
(green top tube) within 30 minutes of collection
Test Time: As little as 4 seconds
Result Value: Plasma calibrated
Assay Method: Electrochemical
Reference Method: Yellow Springs Instrument (YSI)
Power Supply: One 3V lithium battery #CR2032 (non-rechargeable)
Total power when active at full battery = 8.6 mW
Battery Life: Approximately 2146 tests or 1.5 years
Automatic Shut-off: After 2 minutes of non-use
Weight: 1.66 ounces
Size: 3.44” x 2.16” x 0.69”
Memory Size: 500 results
System Operating Range (Meter and Test Strips):
Relative Humidity: 10 - 90% (non-condensing)
Temperature: 50° - 104°F
Hematocrit: 20 - 60%
Note: Use within specified environmental conditions only.
Chemical Composition
TRUEtest Test Strips: Glucose dehydrogenase - PQQ (Acinetobacter
calcoaceticus), mediators, buffers and stabilizers
TRUEtest Glucose Control Solution: Water, d-glucose, buffers,
viscosity enhancing agent, salts, dye and preservatives