Glucose Control Solution Bottle Label
1. Lot Number (LOT) - Used for identification of lot for QC Form, used as a
reference if calling for assistance.
2. Expiration Dates (EXP) - The printed Expiration Date is located on the
Control bottle label next to EXP. Write the date first
opened on bottle label. Discard bottle and unused Control
Solution if either date printed after EXP on Control bottle
label or 3 months after date written on bottle label has
passed.
3. Control Level - Three levels of TRUEtest Glucose Control Solution are
available, Levels 1, 2, and 3. We recommend testing at
least 2 levels of Control. For testing frequency and the
number of Control levels tested, refer to your facility
quality control procedure/policy to test.
TRUEresult
®
Quality Assurance / Quality Control Manual
22
1
2
3