Caring for the TRUEresult System
• Store the System (Meter, Test Strips, Glucose Control Solution) in an
area protected from liquids, dust and dirt.
• Store the System in a dry place at room temperature, 36° - 86°F.
DO NOT REFRIGERATE OR FREEZE.
Meter
• Wipe Meter with a clean, lint-free cloth dampened with 10% bleach
and water, or an OSHA approved disinfectant.
• Never put Meter in liquids or allow any liquids to enter Test Port.
• Do not use 100% Isopropyl alcohol to clean the Meter. It may result
in damage to the Meter Display.
TRUEtest Test Strips
• Store Test Strips in original vial only. Do not transfer old Test Strips
into new vial or store Test Strips outside of vial.
• Write the date first opened on Test Strip vial label. Discard vial and
unused Test Strips if either EXP date printed on Test Strip vial label
or 4 months after date written on vial label has passed.
• Close Test Strip vial immediately after removing one Test Strip. Store
in a dry place at room temperature below 86°F.
DO NOT REFRIGERATE OR FREEZE.
• Do not reuse Test Strips.
• Do not bend, cut or alter Test Strips in any way.
TRUEtest Glucose Control Solution
• Write the date first opened on Control bottle label. Discard bottle if
either EXP date printed on bottle label or 3 months after date written
on bottle label has passed.
• After use, wipe bottle tip clean and recap tightly.
• Store at room temperature, 36° - 86°F.
DO NOT REFRIGERATE OR FREEZE.
TRUEresult
®
Quality Assurance / Quality Control Manual
40