EasyManua.ls Logo

McQuay GZ 035AM

McQuay GZ 035AM
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IOMM ACZ/AGZ-1 87
Table 51, MENU 23 Misc Setup
Screen Display Factory Setpoint Range
1 Unit Type= Air Cooled - -
2 Model AGZ 4 Stages - -
3 # Compressors = 4 - -
4 # Stages = 4 - -
5 Fan Stg/Cir = 2
6 Begin PD = xx PSI 54 42 – 70 psi
7 End PD – xx PSI 34 20 - 40
8 Full Pump DN - NO NO NO - YES
9 EndFullPD = xx PSI 20 psi 10 – 30 psi
10 Low Ambi Opr = NO NO No - YES
11 Low Ambi Temp = xx°° F 60 10 - 100
12 LvgEvapAdj = 0.0°° F -0.8 – +0.8
13 EntEvapAdj = 0.0°° F -0.8 – +0.8
14 #1 EvapAdj = 0.0 psi -13.8 – +13.8
15 #2 EvapAdj = 0.0 psi -13.8 – +13.8
16 #1 CondAdj= 0.0 psi (-13.8 - 13.8)
17 #2 CondAdj= 0.0 psi (-13.8 - 13.8)
18 OAT Select = Lc1 Lc1 NONE-LcL-Rmt
19 OAL Lockout = No NO NO – YES
20 Lockout T = x°° F 0°° F F - 80°° F
21 Alarm = Closed CLOSED CLOSED – OPEN-BLINK(N/O)
22 IDENT = GZ2E01A
Table 52, MENU 24 #1 Curr Alarm
Screen Display
1 Current Alarm
2 @ 0:00 0/00/00
3 Evap = x.x psi (kPa)
4 Cond = x.x psi (kPa)
5 SuctLine=xxx.x °F(°C)
6 LiquisLn=xxx.x °F(°C)
7 Evap Lvg=xxx.x °F(°C)
8 OAT=xx.xx°F(°C)
9 Capacity= xxx%
10 Fan Stage = x
Table 53, MENU 25 #2 Curr Alarm
Screen Display
1 Current Alarm
2 @ 0:00 0/00/00
3 Evap = x.x psi (kPa)
4 Cond = x.x psi (kPa)
5 SuctLine=xxx.x °F(°C)
6 LiquisLn=xxx.x °F(°C)
7 Evap Lvg=xxx.x °F(°C)
8 OAT=xx.xx°F(°C)
9 Capacity= xxx%
10 Fan Stage = x
Table 54, MENU 26 #1 PrevAlarm
Screen Display
1 1 Current Alarm
2 1 x:xx x/xx/xx
3 2 Current Alarm
4 2 x:xx x/xx/xx
5 3 Current Alarm
6 3 x:xx x/xx/xx
7 4 Current Alarm
8 4 x:xx x/xx/xx
9 5 Current Alarm
10 5 x:xx x/xx/xx
Table 55, MENU 27 #2 PrevAlarm
Screen Display
1 1 Current Alarm
2 1 x:xx x/xx/xx
3 2 Current Alarm
4 2 x:xx x/xx/xx
5 3 Current Alarm
6 3 x:xx x/xx/xx
7 4 Current Alarm
8 4 x:xx x/xx/xx
9 5 Current Alarm
10 5 x:xx x/xx/xx
Trouble Analysis for the AGZ MicroTech Controller
Microprocessor Control Board
The Microprocessor Control Board (MCB) is shown in Figure 31. It contains a microprocessor that is
preprogrammed with the software required to monitor and control the chiller. The various MCB
connections and components are described below.

Related product manuals