08.0 Procedures for disposal and precautions
- The device must be disposed of and treated as waste for separate collection;
- The purchaser may return the device at the end of its working life to the dealer supplying a new
device. Instructions for disposal are available from Mectron S.p.A.;
- Failure to comply with the above instructions could lead to application of a penalty in
accordance with Directive 2002/96/EC.
DANGER: Hospital waste.
The following must be treated as hospital waste:
- Inserts, when worn or broken;
- Wrench for tightening the inserts, if worn or broken.
09.0 Inserts for Mectron scaler and
how to use them
S1 - S2 - S3 - S4 - S5: For general use in scaling;
S6 - S7 - S8: High-efficiency, for general use in scaling for removing large quantities of calculus;
P1 - P2 - P3 - P4: For general use in removing concretions from root surfaces;
P10 - P11 - P12 - P13 - P14: For general use in removing concretions from deep root surfaces;
R1 - R2 - R3 - R4 - R5: For general use in retrograde root canal treatment;
D1 - D2 - D3 - D4: For general use in condensing amalgam - burnishing fillings - lateral condensa-
tion of gutta-percha - removal of crowns, bridges and pins;
CM1 - CM2 - CM3 - CM4: For marginal finishing;
ER1 - ER2 - ER3 - ER4 - ER5: For general use in orthograde endodontics;
E1 - E2: Holder for 120° or 90° files, for endo files.
10.0 Trouble-shooting
If the device does not seem to be working correctly, read the instructions again and then check the
following table.
PROBLEM POSSIBLE CAUSE SOLUTION
The device will not switch on.
The blue LED is off (Fig.7
- Rif.G).
The power supply unit is not
properly connected to the
mains outlet or to the device.
The power supply unit and/or
is out of use or ’the internal
fuse is blown.
Make sure that the power
supply unit is properly con-
nected.
Check that the power outlet is
in proper working order.
Contact an authorised
MECTRON Service Centre.
The device is switched on but
will not work.
The pedal is not properly
plugged into its socket.
The pedal is out of order.
Plug the pedal in correctly.
Contact an authorised
MECTRON Service Centre.
A slight hissing noise is made
by the scaler handpiece while
it is working.
The insert is not
properly tightened onto the
scaler handpiece.
Unscrew the insert and screw
it back into place correctly.
One of the following wordings
appears in the display:
WARNING 1
WARNING 2
WARNING 4.
Scaler handpiece not
connected to the device.
Improper functioning of the
tuning circuit.
Handpiece failure.
Check whether the handpiece
is properly connected to its
cord.
Contact an authorised
MECTRON Service Centre.
Contact an authorised
MECTRON Service Centre.
The message WARNING 35
appears on the display for a
few seconds.
The switching off procedure
was not carried out correctly
after the last time it was used.
The device switched off
during use due to an external
interference.
Switch the device off
correctly. See point 05.2.
Consult the tables under point
12.1.