EN
To perform a cleaning cycle of every
handpiece, tube, or connector used during
the same treatment, repeat the operations
described above, from point 5 onward;
IMPORTANT After the cleaning cycle with
the enzymatic detergent solution, perform
a new cleaning cycle with distilled water to
remove detergent residues.
PIEZOSURGERY
®
white
To enter the cleaning mode keep the
PUMP/FLUSH button on the touch
keyboard pressed; the LED of the level 6
RIWKHIXQFWLRQ,UULJDWLRQÁDVKHVDOORWKHU
buttons are disabled. To stop the FLUSH
function release the button;
Go to point 6;
1
0
5a
6
7
2QFH\RXKDYHÀQLVKHGFOHDQLQJWKHLUULJDWLRQFLUFXLWPRYHRQWRWKHFOHDQLQJDQGVWHULOL]DWLRQ
of each single part.