8 Replace canister
WARNINGS
The canister is replaced on the basis of a visual check or according to the instructions on
the Thopaz
+
display (warning signal). Wear gloves for all operations.
CAUTIONS
Canister changes shall be recorded for safety reasons.
34 | Replace canister
2. Clamp off patient tubing with the tube clamp.
4. 4.1
Release the canister by
pressing the release button.
4.2 Remove canister.
1. Prepare sterile canister (with/without solidifier).
5. 5.1 Unpack the canister.
5.2 Position the openings at
the top and attach the
bottom part of the canister
to Thopaz
+
.
6. 6.1 Push the canister into
Thopaz
+
until you hear a
click.
Safety related tip
Using a 2 l canister
refer to chapter 7.3.1 –
Using 2 l canister.
3. 3.1 Switch Thopaz
+
to
Standby by pressing
«Standby» for longer than
3 seconds.
3 sec.