Changing settings | 37
Thopaz
+
is delivered with the following factory settings. These
settings can be changed and saved as the new standard.
Pressure: -2.0 kPa
Choose to max. -10 kPa (-100 cmH
2
O,
-100 mbar, -80 mmHg).
Pressure unit: kPa
Choose between kPa, cmH
2
O, mmHg, mbar.
Measured Hide
pressure: Choose between hide and show (only
relevant in the air leak history).
This feature is added for advanced Thopaz
+
users. A higher measured pressure than set
pressure can be an indication that the patient
is able to regulate the negative pressure in
the inter-pleural space.
Fluid time 1: 24 h
Choose between 24 h / 12 h / 8 h / 6 h
(only relevant in the advanced display).
Shows the total amount of collected fluid
over the last X hours.
Fluid time 2: 1 h
Choose between 3 h / 1 h / 0.5 h (30 min) /
0.25 h (15 min) (only relevant in the advanced
display).
Shows the total amount of collected fluid
over the last X hours/minutes.
Fluid warning: OFF
Choose between OFF / 500 ml / 400 ml /
300 ml / 200 ml / 100 ml / 50 ml
The Fluid warning 407 appears if Thopaz
+
detected a higher fluid collection in the last
5 min than the desired parameter.
Display: Basic
Choose between Basic and Advanced dis-
play.
Language: English
Choose your language.
Save as Choose if you want to save the new
standard: parameter as a standard (guidance see chapter
10.3 – Save a changed parameter as the new
standard).
Factory settings
10.1
10 Changing settings