42 | Troubleshooting and alarms | General cleaning guidelines
Visible on display
Error
number
Problem
description
Troubleshooting
instructions
Remarks / Potential
source of error
Pressure
Internal error
Internal error
XXX
1. Switch Thopaz
+
off
and on again
2. Inform Medela
Customer Service
Safety related tip
The acoustic alarm «Leak in system – 301» is deactivated for approximately 6 minutes
after Thopaz
+
is switched on.
13 General cleaning guidelines
WARNINGS
After each use, the parts that have been in contact with aspirated secretions are to be
cleaned and disinfected or disposed.
General notes
For detailed information refer to Medela cleaning guideline: 200.2391.
This cleaning guideline relates to the cleaning of the entire Thopaz
+
product ranges see
Appendix A.
Never place electrical devices in water or other liquids, specifically Thopaz
+
and the docking
station.
Do not spray or pour liquid directly onto Thopaz
+
/docking station.
– Thopaz
+
product range cannot be sterilized.
Rinsing the Thopaz
+
product range in a washing machine is not permitted.
Immersion disinfection, thermal disinfection and ultrasound cleaning for the Thopaz
+
product
range are not permitted.
Higher temperature than 45°C, 113°F can cause protein coagulation which can lead to
problems further in the process.
Clean all surfaces immediately after use to avoid residues from drying and to prevent growth
of microorganisms.
Do not use cleaning agent/detergent based on phenol.
Do not use steel brushes or steel wool for cleaning.
Store medical products dry and dust free.