Troubleshooting and alarms | 41
Visible on display
Error
number
Problem
description
Troubleshooting
instructions
Remarks / Potential
source of error
Pressure
Warning
306 Canister full 1. Clamp tube
2. Go into Standby
3. Replace canister
4. Continue with «On»
5. Unclamp tubing
Yes
401 Battery low Connect Thopaz
+
to mains
Remaining time of
battery approx. 30 min.
Yes
402
USB connection
not permitted
Unplug USB cable USB connection not
permitted during
operation or in the
vicinity of the patient.
Yes
405 Standby Switch Thopaz
+
on or off In Standby for 5 minutes. No
406 Temperature
too high
Prepare backup device Do not expose Thopaz
+
to heat sources.
Yes
408
Canister change
detected
Please confirm if canister
has been changed
Yes
409 Fluid level
sensor out
of order
– Fluid measurement
unavailable
Confirm and continue
therapy
Yes
Alarm
301 Leak in system Check system for leaks – Disconnection.
Missing seal at
vacuum port.
No
302 Tubing clogged 1. Clamp catheter
2. Clamp tubing
3. Go into Standby
4. Replace tubing
5. Continue with «On»
6. Unclamp catheter
Tubing is kinked or
clogged for approx.
10min.
– Thopaz
+
detects
clotting of Thopaz
+
tubing, NOT of the
patient catheter.
No
305 Battery empty Connect Thopaz
+
to
mains
No
311 Selftest failed Snap the canister out
and
in again
Do not switch Thopaz
+
on if the patient is already
connected.
No
313 Filter clogged 1. Clamp tubing
2. Replace canister
3. Continue with «On»
4. Unclamp tubing
Filter in canister clogged. No
315 Thopaz
+
overheated
Replace Thopaz
+
No
407 Fluid alarm Patient is losing a large
amount of fluid
Yes