Warnings and safety instructions | 9
CAUTIONS
No modification of this equipment is allowed.
Before connecting Thopaz
+
to the power supply, please verify that the supply voltage
corresponds to that given on the device specification plate.
Incorrect use of Thopaz
+
can cause pain and injury to the patient.
Do not use sterile accessories when the sterile packaging is damaged.
Non sterile and reusable accessories must be cleaned and disinfected according to
chapter 13 – General cleaning guidelines.
For the initial startup do not switch Thopaz
+
on before first charging the battery.
The Thopaz
+
2 l canisters are not intended for portable use (not by hand and not using
the carrying strap).
For safety reasons canister changes within the appropriate fluid level shall be recorded.
Before diagnosing controll detected fluid value for plausability.
For patients for whom a breakdown of the device can lead to a critical situation a
replacement device must always be available.
Federal US law restricts this device to sale by or on the order of a health professional.
Safety related tip
Never place Thopaz
+
with an attached 2 l canister on the docking station without
attaching the optional adapter docking station 079.0038 for Thopaz
+
2 l canister first.
The acoustic alarm «Leak in system – 301» is deactivated for approximately 6 minutes
after Thopaz
+
is switched on.
Everytime a new patient is connected, it is recommended to carry out the functional
check.
Before the first application in accordance with the intended use, an inspection according
to IEC 62353 (see Appendix B) is recommended to acquire the reference values.